The Single Use Cytology Brush V BC-V600P-3010 is intended to be used to collect tissue specimens in the biliary system in combination with an endoscope.
BC-V600P-3010 has been designed to collect specimens or cells endoscopically for cytologic examination in conjunction with duodenoscopes. BC-V600P-3010 is inserted into the duodenoscope through an instrument channel to reach the target area, where, by rubbing the brush on the target area, specimens or cells can be collected. BC-V600P-3010 is then withdrawn from the duodenoscope channel with the collected samples.
The handle is connected to operating wires. The grip on the handle portion controls the brush movement inside the tube. By pushing/pulling the handle, the brush at the distal end of the Subject device can be protruded from/drawn into the tube. Cells can be collected by moving over the target site with the brush.
The BC-V600P-3010 consists of:
A. Handle
- Grip: used to push to extend the brush from the tube. Pull to retract the brush back into the tube.
- Operation pipe: used to connect the grip with the wires so that it can transfer the force to push/pull the brush/distal tip.
- Locking cap: used for fixing the location of the brush by turning the locking cap clockwise.
- Injection port: Available to attach a syringe here to inject contract medium.
- Hook: attached to the boot of the endoscope or ET Control Device V (sold separately).
B. Insertion portion
- Wires: connect operation pipe with brush and distal tip to push/pull the brush from/into the tube
- Tube: works as the channel that the contrast media is injected through during use
- V marking: gives the user an approximate indication of how far the instrument has been inserted into the endoscope,
- Radiopaque marks: allow the distal end of the tube to be visualized on the X-ray monitor
- Brush: used to brush the target area to collect samples.
- Distal tip: When using a guidewire, the guidewire is inserted into the distal tip.
N/A
Ask a specific question about this device
(76 days)
Ask a specific question about this device
(203 days)
The Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC631Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope. The preloaded guidewire is used for guiding and exchanging endoscopic accessories for biliary duct, including but not limited to the common bile, cystic, pancreatic and right and left hepatic ducts.
The Single Use Sphincterotome V (Distal Wireguided) KD-VC411Q/VC431Q/VC433Q Series are intended to be used for papillotomy using high-frequency current in combination with an endoscope.
The Single Use 3-Lumen Sphincterotome V KD-V411M/V431M Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope and guidewire.
The DISPOSABLE TRIPLE LUMEN SPHINCTEROTOME KD-411Q/431Q Series is intended to be used for papillotomy using high-frequency current in combination with an endoscope.
The Single Use Sphincterotome KD-VC, KD-V, and KD Series consists of the following devices:
- Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC600 Series (aka CleverCut3V)
- Single Use Sphincterotome V (Distal Wireguided) KD-VC400 Series (aka CleverCut3V)
- Single Use 3-Lumen Sphincterotome V KD-V Series (aka CleverCut3V)
- DISPOSABLE TRIPLE Lumen SPHINCTEROTOME KD Series (aka FlowCut)
The subject devices have separate lumens for guidewire, cutting wire and injection of contrast medium for papillotomy. The knives are pre-curved and have a tapered tip to facilitate insertion into the papilla of Vater. The insertion portion of the KD-V, and KD-VC Series have a V-marking, which the relative insertion length into the endoscope can be confirmed by the positional relationship between the V-marking and the biopsy valve of the endoscope and feature an integrated C-Hook to attach to the endoscope. The distal end of the sphincterotome is coated with CleverCut coating ("Olympus' PFA (perfluoroalkoxy) coating designed to prevent thermal injury to non-target tissue while cutting the papilla of Vater. The KD-VC600 Series offers models that are preloaded with VISIGLIDE guidewires.
The subject devices are provided to the user sterilized by ethylene oxide and intended for single use only.
N/A
Ask a specific question about this device
(245 days)
The Single Use Electrosurgical Knife KD-612 Series is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.
The Single Use Electrosurgical Knife KD-620UR is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.
The Single Use Electrosurgical Knife KD-650 Series is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.
The Single Use Electrosurgical Knife was designed to be used with compatible Olympus endoscopes, A-Cord (power cable), and electrosurgical units to cut tissue using high-frequency current within the digestive tract. The subject device is inserted endoscopically into the body cavity and tissue is cut by applying a high frequency current. The subject device is a sterile (EO) single use electrosurgical knife comprised of a handle and an insertion portion. The handle portion consists of the body and the slider. The insertion portion consists of the distal end with one lumen for the stainless-steel cutting knife and the sheath portion. The slider extends the cutting knife from the distal end of the sheath/tube when pushed and retracts the cutting knife into the sheath/tube when the slider is pulled.
The Single Use Electrosurgical Knife is available in three different models: KD-612, KD-620UR, and KD-650. The primary difference between the models is the tip shape of the distal end.
The subject device has the same technological characteristics and similar design as the applicable predicate device.
N/A
Ask a specific question about this device
(245 days)
The Single Use Suction-Evacuation Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. It is designed to establish a conduit for the treatment of urinary stones during endoscopic procedures.
The Single Use Suction-Evacuation Ureteral Access Sheath is a single-use ureteral access sheath, which can create access to the urinary tract by provide ureteral dilation and a continuous working channel for the introduction of endoscope and instruments during ureteral access procedures. The device can inject fluids and has vacuum pressure design which can facilitate with stone cleaning procedures.
The device consists of Dilator, Sheath, Sheath hub, Sealing cap, Dilator hub and Pressure control vent; Both dilator and sheath outer surfaces are coated with hydrophilic coatings, which provide longlasting lubrication for better expansion and passage. The distal end of the sheath (10cm) can bend with the endoscopes, and the tip is soft which provides seamless transition between sheath core and sheath tube. The dilator and sheath are designed with radiopaque which is visible under X-Ray, can be visualized under X-ray(fluoroscopy) during placement to confirm location.
N/A
Ask a specific question about this device
(260 days)
RO handpiece, AGNES (F) RF handpiece and AGNES (B) RF handpiece of RFMagik are intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
RFMagik is a medical device combined with RF current, to function as an electrosurgical device for use in dermatology and general surgical procedures. It is possible to select and change modes, parameters, outputs, etc. using the panel on the main body. It consists of the main device, LCD screen, two handpieces, single use electrodes, electrode pad, NE pad cable, food switch.
There are three handpieces. RO handpiece, AGNES (F) RF handpiece and AGNES (B) RF handpiece that delivers RF energy through the disposable electrode in the handpiece.
This document is an FDA 510(k) clearance letter for an RF Electrosurgical Device (RFMagik). It states that the device is substantially equivalent to legally marketed predicate devices.
However, the provided document does NOT contain information about acceptance criteria, device performance results, sample sizes, expert ground truth establishment, or clinical study details. The section on "Clinical Testing" explicitly states: "Clinical testing is not a requirement and has not been performed."
The document focuses on:
- Regulatory details: Device classification, product codes, indications for use.
- Technological comparison: Detailed comparison of the subject device (RFMagik) with a primary predicate device (RFMagik Lite) and a reference device (AGNES). This comparison highlights similarities and differences in handpieces, electrodes, output power, etc., and explains why these differences do not affect substantial equivalence.
- Non-clinical testing: Biocompatibility, sterility, shelf-life, and performance bench testing. An "Ex Vivo Study" was conducted on tissue types (liver, skin, muscle) for thermal testing.
Therefore, I cannot fulfill your request for a table of acceptance criteria, device performance, sample sizes for the test set, data provenance, number/qualifications of experts, adjudication methods, MRMC study details, standalone performance, type of ground truth, training set sample size, or how training ground truth was established. This information is typically found in specific study reports or sections of a 510(k) submission that go beyond what is published in the clearance letter itself.
The document indicates that the substantial equivalence was primarily demonstrated through bench testing and comparison to predicate devices, rather than clinical trials or extensive human-in-the-loop performance studies.
Summary of what CANNOT be provided from the given document:
- Table of acceptance criteria and reported device performance: Not present.
- Sample size for the test set and data provenance: No clinical test set. Ex vivo study mentioned, but specific sample sizes are not detailed.
- Number of experts and qualifications for ground truth: Not applicable as no clinical study with expert ground truth review was performed.
- Adjudication method for the test set: Not applicable.
- MRMC comparative effectiveness study: Not conducted.
- Standalone (algorithm only) performance: Not applicable as this is a physical electrosurgical device, not an AI algorithm.
- Type of ground truth used: Not applicable for a clinical study. Ex vivo study used physical tissue, but no "ground truth" akin to medical image labeling.
- Sample size for the training set: Not applicable as there is no mention of an AI/ML algorithm requiring a training set.
- How ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(60 days)
The Single Use Aspiration Needle NA-201SX-4021 is intended to be used for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope.
The Single Use Aspiration Needle NA-201SX-4022 is intended to be used for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope.
The Olympus Single Use Aspiration Needle NA-201SX-4021 and NA-201SX-4022 are designed for use with ultrasound endoscopes for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. These devices are single-use, disposable, sterile (Ethylene Oxide) needles available in two models with different specifications:
• NA-201SX-4021:
- Working Length: 700mm
- Maximum Insertion Portion Diameter: 1.9mm
- Needle Width: 21G
- Normal Needle Length: 20mm
- Maximum Needle Length: 40mm
• NA-201SX-4022:
- Working Length: 700mm
- Maximum Insertion Portion Diameter: 1.8mm
- Needle Width: 22G
- Normal Needle Length: 20mm
- Maximum Needle Length: 40mm
Both models feature an adjustable sheath length and are intended for single use.
This document is a 510(k) clearance letter for a physical medical device (Single Use Aspiration Needle), not an AI/Software-as-a-Medical-Device (SaMD). Therefore, the requested information pertaining to AI/SaMD performance evaluation criteria (such as test set size, data provenance, expert ground truth, MRMC studies, standalone performance, training set details, etc.) is not applicable and is not present in the provided document.
The document discusses the substantial equivalence of the new aspiration needles to a predicate device based on their physical and functional characteristics. The performance data section refers to bench testing of physical attributes, not computational performance or diagnostic accuracy.
Therefore, I cannot provide the requested information as it relates to AI/SaMD. I can, however, extract the acceptance criteria and reported device performance for the physical device based on the provided text.
Acceptance Criteria and Reported Device Performance (Physical Device)
This section summarizes the performance data for the Single Use Aspiration Needle (NA-201SX-4021 and NA-201SX-4022), based on the provided FDA 510(k) summary. These criteria relate to the physical and functional performance of the device, not an AI algorithm.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Test | Acceptance Criteria (Implied/Stated) | Reported Device Performance |
|---|---|---|
| Insertion performance into the endoscope | Smooth insertion and effective tissue penetration | All results met acceptance criteria for smooth insertion and effective tissue penetration |
| Piercing performance of the needle | Effective tissue penetration | All results met acceptance criteria for effective tissue penetration |
| Ultrasound visibility of the needle | Visibility of the needle under ultrasound imaging | All results met acceptance criteria for visibility of the needle under ultrasound imaging |
| Needle extraction and retraction performance | Safe and effective needle extraction and retraction | The device met all predefined criteria for safe and effective needle extraction and retraction |
| Needle aspiration performance | Successful aspiration of target material under test conditions | The device successfully aspirated target material under test conditions |
| Withdrawal performance from the endoscope | Smooth and safe withdrawal from the endoscope | The needle was withdrawn smoothly and safely from the endoscope |
| Needle slider performance | Reliable functioning for needle deployment and retraction | The slider mechanism functioned reliably for needle deployment and retraction |
| Limitation of needle depth | Effective limitation of needle extension to within specified parameters | The device effectively limited needle extension to within specified parameters |
| Needle-to-luer joint pull strength | Withstood required tensile forces without failure | The joint withstood required tensile forces without failure |
| Sheath-to-handle joint pull strength | Connection between sheath and handle remained secure under stress | The connection between sheath and handle remained secure under stress |
| Needle breakage strength | Sufficient structural integrity under load | The needle demonstrated sufficient structural integrity under load |
| Coil separation | Coil remained intact and did not separate under test conditions | The coil remained intact and did not separate under test conditions |
| Biocompatibility (various tests per ISO 10993-1) | Met acceptance criteria for cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility, and aged cytotoxicity | All tests met acceptance criteria, confirming biological safety and biocompatibility |
| Sterilization (Ethylene Oxide) | Achieved a sterility assurance level of 10⁻⁶ and met endotoxin limits | Achieved a sterility assurance level of 10⁻⁶ and met endotoxin limits |
| Shelf-life | Validated a three-year shelf life through accelerated aging and simulated distribution, with packaging integrity and product performance tests passing acceptance criteria | Validated a three-year shelf life, with all packaging integrity and product performance tests passing acceptance criteria |
Regarding the AI/SaMD specific questions:
As stated previously, the provided document is for a physical medical device (Single Use Aspiration Needle), not an AI/SaMD device. Therefore, the following information is not applicable and not provided in the source text:
- Sample size used for the test set and the data provenance: Not applicable. Performance testing was physical bench testing, not data-driven AI evaluation.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's performance is typically established by engineering specifications and direct measurement, not expert review of AI outputs.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This applies to AI-assisted diagnostic or interpretive systems.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This applies to AI algorithms.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" for a physical needle refers to its engineered specifications and expected mechanical/biological performance, verified through bench testing.
- The sample size for the training set: Not applicable. This applies to machine learning models.
- How the ground truth for the training set was established: Not applicable. This applies to machine learning models.
Ask a specific question about this device
(90 days)
The Single Use Cannula V PR-V223Q/V414Q/V416Q/V418Q/V420Q/V427Q/V434Q/V435Q are intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
The Single Use 2-Lumen Cannula V PR-V614M is intended to be used to inject contrast medium in the biliary or pancreatic duct in combination with an endoscope.
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series is comprised of nine (9) sterile, single-use, cannulas designed to inject contrast medium in the biliary or pancreatic duct when used in conjunction with a compatible endoscope.
Each device has two sections: the handle (proximal portion) and the insertion portion. The insertion portion is introduced into the biliary or pancreatic ducts through an endoscope. The distal end of the insertion portion is designed for smooth cannulation of the papilla of Vater or the minor papilla. All models are visible under fluoroscopy and feature a distal marking system.
The Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series models are to be used with compatible endoscopes.
The provided 510(k) Premarket Notification document describes a medical device, the "Single Use Cannula V and the Single Use 2-Lumen Cannula V PR Series," and its comparison to a predicate device for demonstrating substantial equivalence.
However, the document does not contain information about acceptance criteria or a study proving the device meets specific acceptance criteria in the context of device performance as typically expected for software-enabled devices or those with diagnostic capabilities.
This submission is for a physical medical device (cannulas) and focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing of its physical properties and biocompatibility. Therefore, many of the requested categories (like sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types) are not applicable to the information provided.
Based on the document, here's what can be extracted and what cannot:
1. Table of Acceptance Criteria and Reported Device Performance:
The document outlines performance data that was provided to demonstrate substantial equivalence, rather than specific and quantitative acceptance criteria with reported numerical device performance against those criteria. The "Analysis" column in the comparison table broadly states "Substantially equivalent" or "Identical," but doesn't provide the detailed numbers that would typically be associated with acceptance criteria for a diagnostic or algorithmic device.
| Performance Data Category | Description |
|---|---|
| Biocompatibility | Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity |
| Sterilization Validation | Per ISO 11135:2014 |
| Ethylene Oxide Residuals | Per ISO 10993-7:2008 |
| Packaging Validation & Shelf Life | Per ISO 11607-1:2019 and ASTM F1980-21 |
| Mechanical Testing & Comparative Testing | Insertion force/Withdrawal force, Insertion w/ Stylet, Attachment and detachment of the hook, Contrast medium infusion, Connection strength, Visibility |
| Human Factors Testing | Verification of device performance |
Acceptance Criteria and Reported Performance (General statement from the document):
The document states: "Non-clinical testing demonstrates that the slight differences in device design do not alter the safety, efficacy, or performance of the subject devices when compared to the predicate devices." and "The non-clinical data demonstrate that the subject device is as safe, as effective, and performs as well as or better than the identified predicate device." This is a qualitative conclusion of meeting equivalence rather than presenting specific numerical acceptance criteria.
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: Not specified in the provided document. The performance data refers to various non-clinical tests, and the sample size for these individual tests (e.g., number of cannulas tested for insertion force) is not detailed.
- Data Provenance: Not applicable in the context of clinical data. The tests are non-clinical (laboratory/bench testing).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Ground truth as typically understood for diagnostic performance (e.g., disease presence/absence) is not relevant for these non-clinical, physical device performance tests. "Human Factors Testing" is mentioned, which would involve experts, but the number and qualifications are not provided, nor is it the type of "ground truth" establishment usually refers to in the context of diagnostic AI.
4. Adjudication method for the test set:
- Not applicable. This is typically used for clinical study endpoints or image interpretation, not for physical performance tests of a cannula.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a physical cannula, not an AI or diagnostic imaging system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable in the conventional sense. The "ground truth" for the non-clinical tests would be the physical properties and functional performance measured against predefined specifications or predicate device performance.
8. The sample size for the training set:
- Not applicable for a physical device where "training set" doesn't apply in the context of machine learning.
9. How the ground truth for the training set was established:
- Not applicable for a physical device.
Ask a specific question about this device
(97 days)
The Single Use Retrieval Basket V FG-V421PR/FG-V422PR/FG-V431P/FG-V432P are intended to be used to retrieve stones from the biliary tract in combination with an endoscope.
The Single Use Retrieval Basket V Series is comprised of four (4), sterile, single-use, retrieval baskets designed to retrieve stones from the biliary tract. The retrieval baskets, manufactured from stainless steel, are provided as an 8-wire or 4-wire type.
Each device has two sections: the handle and the insertion. The grip of the handle is used to control and operate the retrieval basket. The insertion, consisting of the sheath and the retrieval basket, is introduced into the biliary tract through an endoscope.
The subject devices are intended to be used with Olympus endoscopes featuring a working length of less than 1400 mm (model: TJF) and a channel inner diameter of Ø 4.2 mm. The Olympus Lithotriptor (model BML-110A-1 or BML-610A) may be used in case of an emergency.
Legally marketed Olympus guidewires (outer diameter Ø 0.89 mm) may be used with the Single Use Retrieval Basket V FG-V431P and FG-V432P models.
This document, a 510(k) Premarket Notification for the Olympus Single Use Retrieval Basket V (K243807), details the device's substantial equivalence to a predicate device. However, it does not contain the specific information required to address your request regarding acceptance criteria and a study proving the device meets those criteria, particularly in the context of an AI/human comparative effectiveness study.
The provided text focuses on the device's mechanical, material, and operational equivalence to a previously cleared device. It outlines:
- Device Description: Single-use retrieval baskets for stones in the biliary tract.
- Performance Data: Biocompatibility testing (ISO 10993-1), sterilization validation (ISO 11135), packaging validation (ISO 11607-1, ASTM F1980-21), and mechanical testing (Insertion/Withdrawal, Open/Close Basket, Dimensional verification, Grasping Basket Effective Test, Attachment/Detachment of Hook, Injecting Fluid, Strength of Junction).
- Conclusion: The device is substantially equivalent to the predicate based on these non-clinical tests.
Crucially, the document explicitly states:
- "Animal study data and clinical study data were not required to demonstrate substantial equivalence."
This indicates that a study involving human subjects or AI-assisted performance, which would typically contain the data you're requesting, was not part of this 510(k) submission. The acceptance criteria described are for the physical and material properties of the basket, not for diagnostic performance or human-AI interaction.
Therefore, I cannot provide the requested information from this document. The device in question is a medical instrument (a retrieval basket), not an AI/software as a medical device (SaMD) that would have diagnostic performance metrics, ground truth establishment, or multi-reader studies.